Matrices listed below were spiked with certain level of CD55 and the recovery rates were calculated by comparing the measured value to the expected amount of CD55 in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
85-105
95
EDTA plasma(n=5)
92-104
99
heparin plasma(n=5)
88-105
96
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of CD55 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
CD55/DAF(Complement decay-accelerating factor)/CR/TC/CD55 antigen/decay accelerating factor for complement
Notes
For Research Use Only
UniProt ID
P08174
Antigen
CD55
Applications
ELISA
Applications Notes
CD55 ELISA Kit allows for the in vitro quantitative determination of CD55 concentrations in serum, plasma, tissue homogenates and other biological fluids.
Precision
Intra-Assay: CV<8% Inter-Assay: CV<10%
Storage
2-8 °C for 6 months