Rat FGF21 ELISA Kit PicoKine® (96 Tests). Quantitate Rat FGF21 in cell culture supeatants, serum and plasma (heparin, EDTA). Sensitivity: 10pg/ml . The brand Picokine indicates this is a premium quality ELISA kit. Each Picokine kit delivers precise quantification, high sensitivity, and excellent reproducibility. Only our most reliable and effective kits qualify as Picokine, guaranteeing top-tier results for your assays.
Alternative Names
Fibroblast growth factor
Immunogen
Expression system for standard: E.coli; Immunogen sequence: Y30-S208
There is no detectable cross-reactivity with other relevant proteins.
Specificity
Natural and recombinant rat FGF21
Sample Type
Cell culture supeatants, Serum, Plasma
Anticoagulant
heparin or EDTA
Precision
Intra-Assay Precision
Inter-Assay Precision
Sample
1
2
3
1
2
3
n
16
16
16
24
24
24
Mean (pg/ml)
50
155
638
57
172
749
Standard deviation
2.10
6.67
30.62
2.79
11.01
58.42
CV (%)
4.2%
4.3%
4.8%
4.9%
6.4%
7.8%
Background
Fibroblast growth factor 21 is a protein that in humans is encoded by the FGF21 gene. By genomic sequence analysis, Nishimura et al. (2000) identified the FGF21 gene within the 5-prime flanking region of the FUT1 gene on chromosome 19. The protein encoded by this gene is a member of the fibroblast growth factor (FGF) family. FGF family members possess broad mitogenic and cell survival activities and are involved in a variety of biological processes including embryonic development, cell growth, morphogenesis, tissue repair, tumor growth and invasion. FGF21 stimulates glucose uptake in adipocytes but not in other cell types. This effect is additive to the activity of insulin. FGF21 treatment of adipocytes is associated with phosphorylation of FRS2, a protein linking FGF receptors to the Ras/MAP kinase pathway. FGF21 injection in ob/ob mice results in an increase in Glut1 in adipose tissue. FGF21 also protects animals from diet-induced obesity when overexpressed in transgenic mice and lowers blood glucose and triglyceride levels when administered to diabetic rodents.
Reproducibility
Lots
Lot 1 (pg/ml)
Lot 2 (pg/ml)
Lot 3 (pg/ml)
Lot 4 (pg/ml)
Mean (pg/ml)
Standard Deviation
CV (%)
Sample 1
50
48
42
44
46
3.16
6.8%
Sample 2
155
181
179
185
175
11.74
6.7%
Sample 3
638
762
636
688
681
51.19
7.5%
*number of samples for each test n=16.
Notes
Material Required But Not Provided: Microplate Reader capable of reading absorbance at 450nm. Incubator. Automated plate washer . Pipettes and pipette tips capable of precisely dispensing 0.5 µl through 1 ml volumes of aqueous solutions. Multichannel pipettes are recommended for large amount of samples. Deionized or distilled water. 500ml graduated cylinders. Test tubes for dilution.
Store at 4°C for 6 months, at -20°C for 12 months. Avoid multiple freeze-thaw cycles (Ships with gel ice, can store for up to 3 days in room temperature. Freeze upon receiving.)