Mouse PPBP/CXCL7 ELISA Kit PicoKine® (96 Tests). Quantitate Mouse Ppbp in cell culture supeatants, serum and plasma (heparin, EDTA). Sensitivity: 2pg/ml. The brand Picokine indicates this is a premium quality ELISA kit. Each Picokine kit delivers precise quantification, high sensitivity, and excellent reproducibility. Only our most reliable and effective kits qualify as Picokine, guaranteeing top-tier results for your assays.
Alternative Names
Chemokine (C-X-C motif) ligand 7; isoform CRA_b
Immunogen
Expression system for standard: E.coli; Immunogen sequence: K40-Y113
There is no detectable cross-reactivity with other relevant proteins.
Specificity
Natural and recombinant mouse CXCL7
Sample Type
Cell culture supeatants, Serum, Plasma
Anticoagulant
heparin or EDTA
Precision
Intra-Assay Precision
Inter-Assay Precision
Sample
1
2
3
1
2
3
n
16
16
16
24
24
24
Mean (pg/ml)
37
173
408
33
166
406
Standard deviation
1.96
11.76
16.72
2.17
11.62
17.05
CV (%)
5.3%
6.8%
4.1%
6.6%
7%
4.2%
Background
Chemokine (C-X-C motif) ligand 7 (CXCL7), also called PPBP or SCYB7, is a human gene. The encoded protein, Chemokine (C-X-C motif) ligand is a small cytokine belonging to the CXC chemokine family. CXCL7 is mapped to 4q13.3. It is a protein that is released in large amounts from platelets following their activation. CXCL7 is the precursor of the 2 platelet alpha-granule proteins, platelet basic protein (PBP) and connective tissue-activating peptide III (CTAP3). It stimulates various processes including mitogenesis, synthesis of extracellular matrix, glucose metabolism and synthesis of plasminogen activator. It also stimulates the formation and secretion of plasminogen activator by synovial cells.
Reproducibility
Lots
Lot 1 (pg/ml)
Lot 2 (pg/ml)
Lot 3 (pg/ml)
Lot 4 (pg/ml)
Mean (pg/ml)
Standard Deviation
CV (%)
Sample 1
37
37
35
32
35
2.04
5.8%
Sample 2
173
164
153
159
162
7.32
4.5%
Sample 3
408
466
453
448
443
21.66
4.8%
*number of samples for each test n=16.
Notes
Material Required But Not Provided: Microplate Reader capable of reading absorbance at 450nm. Incubator. Automated plate washer . Pipettes and pipette tips capable of precisely dispensing 0.5 µl through 1 ml volumes of aqueous solutions. Multichannel pipettes are recommended for large amount of samples. Deionized or distilled water. 500ml graduated cylinders. Test tubes for dilution.
Store at 4°C for 6 months, at -20°C for 12 months. Avoid multiple freeze-thaw cycles (Ships with gel ice, can store for up to 3 days in room temperature. Freeze upon receiving.)