Matrices listed below were spiked with certain level of SYNPO and the recovery rates were calculated by comparing the measured value to the expected amount of SYNPO in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
94-103
99
EDTA plasma(n=5)
91-105
98
heparin plasma(n=5)
88-97
92
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of SYNPO and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
SYNPO ELISA Kit allows for the in vitro quantitative determination of SYNPO concentrations in serum, plasma, tissue homogenates and other biological fluids.