Matrices listed below were spiked with certain level of PLN and the recovery rates were calculated by comparing the measured value to the expected amount of PLN in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
85-102
96
EDTA plasma(n=5)
87-100
92
heparin plasma(n=5)
85-99
90
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of PLN and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
PLN ELISA Kit allows for the in vitro quantitative determination of PLN concentrations in serum, plasma, tissue homogenates and other biological fluids.
Precision
Intra-Assay: CV<8% Inter-Assay: CV<10%
Storage
2-8 °C for 6 months