Human PLA2G7/Lp-PLA2 ELISA Kit PicoKine® (96 Tests). Quantitate Human PLA2G7 in cell culture supeatants, serum and plasma (heparin, EDTA). Sensitivity: 20pg/ml. The brand Picokine indicates this is a premium quality ELISA kit. Each Picokine kit delivers precise quantification, high sensitivity, and excellent reproducibility. Only our most reliable and effective kits qualify as Picokine, guaranteeing top-tier results for your assays.
There is no detectable cross-reactivity with other relevant proteins.
Specificity
Natural and recombinant human PLA2G7
Subcellular Localization
Secreted, extracellular space.
Sample Type
Cell culture supeatants, Serum, Plasma
Anticoagulant
heparin or EDTA
Precision
Intra-Assay Precision
Inter-Assay Precision
Sample
1
2
3
1
2
3
n
16
16
16
24
24
24
Mean (pg/ml)
1179
7607
2212395
1239
6920
22984
Standard deviation
55.41
418.38
1124.79
75.57
456.72
1539.92
CV (%)
4.7%
5.5%
4.9%
6.1%
6.6%
6.7%
Protein Function
Modulates the action of platelet-activating factor (PAF) by hydrolyzing the sn-2 ester bond to yield the biologically inactive lyso-PAF. Has a specificity for substrates with a short residue at the sn-2 position. It is inactive against long-chain phospholipids.
Background
Lipoprotein-associated phospholipase A2 (Lp-PLA2) also known as platelet-activating factor acetylhydrolase (PAF-AH) is a phospholipase A2 enzyme that in humans is encoded by the PLA2G7 gene. Lp-PLA2 is a 45-kDa protein of 441 amino acids. It is one of several PAF acetylhydrolases. The protein encoded by this gene is a secreted enzyme that catalyzes the degradation of platelet-activating factor to biologically inactive products. Defects in this gene are a cause of platelet-activating factor acetylhydrolase deficiency. Two transcript variants encoding the same protein have been found for this gene.
Reproducibility
Lots
Lot 1 (pg/ml)
Lot 2 (pg/ml)
Lot 3 (pg/ml)
Lot 4 (pg/ml)
Mean (pg/ml)
Standard Deviation
CV (%)
Sample 1
1179
1334
1212
1128
1213
75.86
6.2%
Sample 2
7607
7415
7511
7063
7399
205.52
2.7%
Sample 3
22955
22970
21188
24226
22834
1081.67
4.7%
*number of samples for each test n=16.
Notes
Material Required But Not Provided: Microplate Reader capable of reading absorbance at 450nm. Incubator. Automated plate washer . Pipettes and pipette tips capable of precisely dispensing 0.5 µl through 1 ml volumes of aqueous solutions. Multichannel pipettes are recommended for large amount of samples. Deionized or distilled water. 500ml graduated cylinders. Test tubes for dilution.
Store at 4°C for 6 months, at -20°C for 12 months. Avoid multiple freeze-thaw cycles (Ships with gel ice, can store for up to 3 days in room temperature. Freeze upon receiving.)
Reviews of Human PLA2G7/Lp-PLA2 ELISA Kit PicoKine®