Matrices listed below were spiked with certain level of HO1 and the recovery rates were calculated by comparing the measured value to the expected amount of HO1 in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
85-105
97
EDTA plasma(n=5)
86-100
92
heparin plasma(n=5)
87-105
96
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of HO1 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
HO1 ELISA Kit allows for the in vitro quantitative determination of HO1 concentrations in serum, plasma, tissue homogenates and other biological fluids.