Add a certain amount of COVID-19 S1 into the sample. Calculate the recovery by comparing the measured value with the expected amount of COVID-19 S1 in the sample.
Sample Type
Recovery Range(%)
Average(%)
serum(n=10)
92-104
98
EDTA plasma(n=10)
85-101
95
Heparin plasma(n=10)
87-105
94
Linearity
Dilute the sample with a certain amount of COVID-19 S1 at 1:2, 1:4 and 1:8 to get the recovery range.
Sample Type
1:2
1:4
1:8
serum(n=10)
95-105%
98-104%
89-100%
EDTA plasma(n=10)
82-95%
85-94%
82-96%
Heparin plasma(n=10)
91-98%
84-100%
86-97%
Species
Coronavirus
Specificity
Specifically binds with COVID-19 S1 , no obvious cross reaction with other analogues.
Standard Curve
This product is detected by QC department and meets performance required in the manual. (Laboratory Humidity: 20%-60%; Temperature: 18°C -25°C; Equilibrate TMB substrate to 37°C before staining. After adding into the ELISA wells, incubate for 15min at 37°C in dark.) Due to different assay environments and operations, assay data below and standard curve are provided for reference. Experimenters should establish standard curve according to their own assay.
STD.(pg/ml)
OD-1
OD-2
Average
0
0.109
0.114
0.111
1.562
0.162
0.17
0.165
3.125
0.198
0.208
0.202
6.25
0.31
0.325
0.316
12.5
0.504
0.529
0.514
25
0.865
0.909
0.883
50
1.395
1.466
1.423
100
2.13
2.238
2.173
Sample Type
Other liquid samples, Serum, Cell culture supernatants, Cell lysate, Plasma
Precision
Intra-assay Precision: samples with low, medium and high concentration are tested 20 times on the same plate. Inter-assay Precision: samples with low, medium and high concentration are tested 20 times on three different plates.
Item
Intra-assay Precision
Inter-assay Precision
Sample
1
2
3
1
2
3
n
20
20
20
20
20
20
Mean (pg/ml)
3.24
13.14
48.7
3.05
12.72
50.01
Standard deviation
0.16
0.64
2.35
0.14
0.69
2.54
CV(%)
4.95
4.89
4.82
4.69
5.43
5.08
Stability
Perform the stability test for the sealed kit at 37°C and 2-8°C and get relevant data.
ELISA kit(n=5)
37°C for 1 month
2-8°C for 6 months
Average(%)
80
95-100
Notes
For Research Use Only
Usage Notes
Step 1: Take out the required plate wells, add 50ul Cap/Det Ab into each well, then add 50ul Standard or Sample into individual well. (When adding standard or sample, the disposable tip lightly touches the liquid level. Change the disposable tips for different samples and standards.) Gently tap the plate for 10s to ensure thorough mixing then static incubate for 60 minutes at 37°C.
Washing: Wash the plate twice without immersion.
Step 2: Add 100ul HRP-Streptavidin (orange) into each well, seal the plate and static incubate for 30 minutes at 37°C.
Washing: Wash the plate five times without immersion.
Step 3: Add 90ul TMB substrate solution, seal the plate and static incubate for 10-20 minutes at 37°C. (Accurate TMB visualization control is required.)
Step 4: Add 50ul stop solution. Read at 450nm immediately and calculate.
Antigen
S1
Applications
ELISA
Applications Notes
COVID-19 Spike protein ELISA Kit allows for the in vitro quantitative determination of COVID-19 S1 concentrations in serum, plasma, tissue homogenates and other biological fluids.