This kit is based on Double antigen-Sandwich ELISA detection method and takes 4h assay time. The microplate provided in this kit has been precoated with antigen. Add standard and properly diluted sample into relevant well respectively. After incubation, wash unbound components. Add biotinylated detection antigen. Then, it binds with COVID-19 Neutralizing Antibody bound to precoated antigen. Wash unbound components and add HRP-Streptavidin Conjugate (SABC). Wash unbound components again and add TMB substrate solution. Then, TMB was catalyzed by HRP to produce a blue color product that turned yellow after adding a stop solution. Read the O.D. absorbance at 450nm in a microplate reader. Calculate the concentration of COVID-19 Neutralizing Antibody in the sample by plotting standard curve. The concentration of the target substance is proportional to the OD450 value.
Alternative Names
Neutralizing antibody ELISA Kit
Detection Method
Double Antigen
Recovery
Add a certain amount of COVID-19 Neutralizing Antibody into the sample. Calculate the recovery by comparing the measured value with the expected amount of COVID-19 Neutralizing Antibody in the sample.
Sample Type
Recovery Range(%)
Average(%)
serum(n=10)
87-104
94
EDTA plasma(n=10)
86-105
99
Heparin plasma(n=10)
86-105
92
Linearity
Dilute the sample with a certain amount of COVID-19 Neutralizing Antibody at 1:2, 1:4 and 1:8 to get the recovery range.
Sample Type
1:2
1:4
1:8
serum(n=10)
86-97%
86-98%
80-93%
EDTA plasma(n=10)
87-105%
85-100%
81-97%
Heparin plasma(n=10)
85-103%
86-100%
81-100%
Species
Universal
Specificity
Specifically binds with COVID-19 Neutralizing Antibody , no obvious cross reaction with other analogues.
Standard Curve
This product is detected by QC department and meets performance required in the manual. (Laboratory Humidity: 20%-60%; Temperature: 18°C -25°C; Equilibrate TMB substrate to 37°C before staining. After adding into the ELISA wells, incubate for 15min at 37°C in dark.) Due to different assay environments and operations, assay data below and standard curve are provided for reference. Experimenters should establish standard curve according to their own assay.
Intra-assay Precision: samples with low, medium and high concentration are tested 20 times on the same plate. Inter-assay Precision: samples with low, medium and high concentration are tested 20 times on three different plates.
Item
Intra-assay Precision
Inter-assay Precision
Sample
1
2
3
1
2
3
n
20
20
20
20
20
20
Mean (ng/ml)
3.02
12.5
48.48
3.04
12.92
49.79
Standard deviation
0.15
0.62
2.48
0.16
0.63
2.48
CV(%)
4.82
4.98
5.12
5.23
4.88
4.98
Stability
Perform the stability test for the sealed kit at 37°C and 2-8°C and get relevant data.
ELISA kit(n=5)
37°C for 1 month
2-8°C for 6 months
Average(%)
80
95-100
Notes
For Research Use Only
Usage Notes
Step 1: Take out the required plate wells, add 50ul Cap/Det Ab into each well, then add 50ul Standard or Sample into individual well. (When adding standard or sample, the disposable tip lightly touches the liquid level. Change the disposable tips for different samples and standards.) Gently tap the plate for 10s to ensure thorough mixing then static incubate for 60 minutes at 37°C.
Washing: Wash the plate twice without immersion.
Step 2: Add 100ul HRP-Streptavidin (orange) into each well, seal the plate and static incubate for 30 minutes at 37°C.
Washing: Wash the plate five times without immersion.
Step 3: Add 90ul TMB substrate solution, seal the plate and static incubate for 10-20 minutes at 37°C. (Accurate TMB visualization control is required.)
Step 4: Add 50ul stop solution. Read at 450nm immediately and calculate.
Antigen
COVID-19
Applications
ELISA
Applications Notes
COVID-19 Neutralizing Antibody ELISA Kit allows for the in vitro quantitative determination of COVID-19 Neutralizing Antibody concentrations in serum, plasma, tissue homogenates and other biological fluids.
Storage
2-8°C(Sealed), Don't cryopreserve.
Reviews of COVID-19 Neutralizing Antibody ELISA Kit